TECHNICAL DATA

Battery power supply Lithium Polymer (LiPo) battery 3.400 mAh
Lifetime Lifetime 8 hours (min@max load), 20 hours (max)
AC Power supply 230 Vac with standard USB battery charger
Status LEDs power supply
error
datalogger on
battery status
Insulation Battery-powered instrument, intrinsically isolated
Overvoltage protection 230 Vac max without permanent damages (TBC)
Rejection 50/60 Hz
Sampling rate 10 Hz
Operating modes Signal measurement
Signal generation (with ramps)
Data logging
Buzzer for overload reporting an inability to simulate the requested value
Standard bushings Nr. 4 test leads 4mm
Thermocouple connection Mini plug (7,9 mm) for TC signal measurement or generation
MicroUSB functions power supply
FW upgrade
ModBUS communication (virtual COM)
Ethernet communication Nr. 1 100Mbit Ethernet port
Operating temperature -20..50°C  (without charging), 0-45°C during charge
Error display Overvoltage (V > 28 V)
Under voltage (V < -0,2 V)
Over-current (I > 24 mA)
Under current (I < -0,1 mA)
Resistance < 18 Ohm
Resistance > 400 Ohm
Load cell fail supply
Blink value (voltage/current/resistance/bridge generation failure)
Humidity  30..90 % non condensing
Dimension (WxHxD) 88 x 147 x 25 mm
Weight 330 g
Protection degree IP20
Equipment Connection cables (4), micro USB/USB cable, battery charger
Calibration certificate On-demand
Standard CE
Norms EN61326-1; EN61010-1
SIGNAL MEASUREMENT  
Current 0..24 mA active & passive; protection ± 28 V 
Voltage (V) 0,0÷27 V
Voltage -10mV÷+90mV
Thermocouple Type J,K ,T, E, N, R, S, B, L
Thermoresistors (2,3,4 wires) Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120
Load cell 350 Ohm; -0,2÷+2,4mV/V
Pulse max count 1.000 Hz
Frequency signal 0,1..1.000 Hz
Accuracy 0,03%; 0,04% for current
Resolution 1 µA; 1 mV; 5 µV;  0,1°C; 0,1uV/V 
SIGNAL GENERATION  
Current 0,1..24 mA active & passive; protection ± 28 V 
Voltage (V) 0,1÷24 V
Voltage (mV) -10mV÷+90mV
Thermocouple Tipo J,K ,T, E, N, R,S, B, L
Thermoresistors (2 wires) Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120
Load cell 350 Ohm; -0,2..+2,4mV/V
Pulse min 0,5 ms (1..24V) pulses number settable
Frequency signal 0,1..1.000 Hz
Accuracy 0,03%; 0,04% for current
Resolution 1 µA; 1 mV; 5µV;  0,1°C; 0,02 Ohm;  0,1 uV/V;
RAMP FUNCTIONS  
Signals Current/Voltage/TC/RTD/Load cell
Functions single or in the Loop
Type MAX 9 segments, ramp resolution 100 ms, minimum ramp 1 second
DATA LOGGING  
Datalogger Yes
Sampling rate >500 ms

Handbücher, Flyer

 
IT EN DE
Multifunction Calibrators
pdf icon it pdf icon en -
User Manual
pdf icon it pdf icon en pdf icon de
Manual of ModBUS registers
pdf icon it pdf icon en -

Tools

 
LabVIEW driver
zip icon Multi

Zertifizierungen

 
FAC-SIMILE Calibration report
pdf icon Multi
CE Declaration of Conformity
pdf icon Multi
REACH and ROHS declarations
pdf icon Multi
Customs codes
pdf icon Multi
0
0
0

SENECA offices will be closed for Christmas holidays from 24 December to 6 January. Business activities will resume as normal on Monday, 7 January. Orders received from 20 December will be dispatched from Monday, 7 January. We thank you for your cooperation and wish you happy holidays!

Produktversionen

MSC-D

Multi-signal and multi-function portable smart calibrator with display

Verfügbarkeit zu bestätigen, kontaktieren Sie uns ?

Zubehörs

CU-A-MICROB

USB cable plug USB-A - MicroUSB-B - 5P ( KIT - USB , MY2 , Z109REGBP )

Verfügbarkeit zu bestätigen, kontaktieren Sie uns ?
CE-RJ45-RJ45-R

Ethernet cable (right) ( RJ45 / RJ45 )

In begrenzter Menge verfügbar ?