The USB-ISO accessory is mandatory for certain use cases:
- in case you need to connect several devices (e.g. several MSC calibrators) to the same workstation (a USB-ISO isolator is required for each device)
- in case your MSC calibrator needs to be connected to a grounded circuit
- in case several MSC calibrators are to be connected to the same circuit at the same time
TECHNICAL DATA
Battery power supply | Lithium Polymer (LiPo) battery 3.400 mAh |
Lifetime | Lifetime 8 hours (min@max load), 20 hours (max) |
AC Power supply | 230 Vac with standard USB battery charger |
Status LEDs | power supply communication error pairing BT datalogger on (TBD) battery status |
Insulation | Battery powered instrument, intrinsically isolated |
Overvoltage protection | 230 Vac max without permanent damages (TBC) |
Rejection | 50/60 Hz |
Sampling rate | 10 Hz |
Operating modes | Signal measurement Signal generation (with ramps) Datalogging |
Buzzer | for overload reporting and inability to simulate the requested value |
Standard bushings | Nr. 4 test leads 4mm |
Thermocouple connection | Mini plug (7,9 mm) for TC signal measurement or generation |
MicroUSB functions | power supply FW upgrade ModBUS communication (virtual COM) |
Wireless communication | Bluetooth Low Energy 4.1 for iOS or Android device connection |
Mobile app | MSC by SENECA. Language available: ITA, ENG, DEU, FRA, ESP |
Operating temperature | -20..50°C (without charging), 0-45°C during charge |
Error display | Overvoltage (V > 28 V) Under voltage (V < -0,2 V) Over current (I > 24 mA) Under current (I < -0,1 mA) Resistance < 18 Ohm Resistance > 400 Ohm Load cell fail supply Blink value (voltage/current/resistance/bridge generation failure) |
Humidity | 30..90 % non condensing |
Dimension (WxHxD) | 88 x 147 x 25 mm |
Weight | 330 g |
Protection degree | IP20 |
Equipment | Connection cables (4), microUSB/USB cable, battery charger |
Calibration certificate | On demand |
Standard | CE |
Norms | EN61326-1; EN61010-1 |
SIGNAL MEASUREMENT | |
Current | 0..24 mA active & passive; protection ± 28 V |
Voltage (V) | 0,0÷27 V |
Voltage | -10mV÷+90mV |
Thermocouple | Type J,K ,T, E, N, R, S, B, L |
Thermoresistors (2,3,4 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
Load cell | 350 Ohm; -0,2÷+2,4mV/V |
Pulse | max count 1.000 Hz |
Frequency signal | 0,1..1.000 Hz |
Accuracy | 0,03%; 0,04% for current |
Resolution | 1 µA; 1 mV; 5 µV; 0,1°C; 0,1uV/V |
SIGNAL GENERATION | |
Current | 0,1..24 mA active & passive; protection ± 28 V |
Voltage (V) | 0,1÷24 V |
Voltage (mV) | -10mV÷+90mV |
Thermocouple | Tipo J,K ,T, E, N, R,S, B, L |
Thermoresistors (2 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
Load cell | 350 Ohm; -0,2..+2,4mV/V |
Pulse | min 0,5 ms (1..24V) pulses number settable |
Frequency signal | 0,1..1.000 Hz |
Accuracy | 0,03%; 0,04% for current |
Resolution | 1 µA; 1 mV; 5µV; 0,1°C; 0,02 Ohm; 0,1 uV/V; |
RAMP FUNCTIONS | |
Signals | Current/Voltage/TC/RTD/Load cell |
Functions | single or in Loop |
Type | MAX 9 segments, ramp resolution 100 ms, minimum ramp 1 second |
DATALOGGING | |
Datalogger | Yes |
Sampling rate | >500 ms |