TECHNICAL DATA
Power supply | via Micro USB port |
Battery | 1 Lithium Polymer (LiPo) 3400 mAh battery |
Startup | Button or automatic (via Ethernet for testing machines) |
Autonomy | 4 hours (minimum at maximum load) |
Isolation | A battery-powered instrument with intrinsic isolation |
Degree of protection | IP20 |
Overvoltage protection | 240 Vac max without permanent damage |
Operating temperature | 0..35°C |
Storage temperature | -20..50°C (not charging), 0..40°C during charging |
Humidity | 30..90 % non condensing |
Dimensions (WxHxD) | 96 x 171 x 44 mm |
Weight | 395 g |
Rejection | 50/60 Hz |
Sampling Frequency | 10 Hz |
Operating Modes | Meter, Generator, Datalogger, Ramps |
Display | 5-point capacitive touch display, TFT-LCD 262k colors, 3.5”, 53x71mm, 320x480 resolution, with backlighting |
Management systems | Multilingual touch interface (Italian, English, German, French, Spanish) Web app via HTML5 browser |
Weight | 330 g |
Approval | CE |
Standards | EN61326-1; EN61010-1 |
MEASURING ACCURACY | |
Accuracy | 0.03% basic, 0.04% for current |
Resolution | 1 µA; 1 mV; 5 µV; 0.1°C; 0.1uV/V |
GENERATION ACCURACY | |
Accuracy | 0.03% basic, 0.04% for current |
Resolution | 1 µA; 1 mV; 5µV; 0.1°C; 0.02 Ohm; 0.1 uV/V; |
INTERFACES AND SIGNALS | |
Buttons | On / Off |
LEDs | Power on LED Battery charge status LED Datalogger on LED |
Buzzer | Buzzer to signal overload and impossibility to simulate the requested value |
Standard bushes | No. 4 bushes 4mm |
Thermocouple connection | Mini plug (7.9 mm) for measuring and simulating thermocouple |
COMMUNICATION | |
Wireless | Integrated 2.4 GHz WiFi (802.11 b/g/n, up to 150 Mbps); Access Point or Station mode |
Ethernet | 1 Fast Ethernet 100 Tx port, ModBUS TCP-IP protocol; max 8 TCP-IP clients, control web server |
MEASURING FUNCTIONS | |
Current | 0.1..24 mA active and passive; protection ± 28 V |
Voltage (V) | 0..26 V |
Voltage (mV) | -10mV..+90mV |
Thermocouple | Type J,K ,T, E, N, R, S, B, L |
Thermocouples (2,3,4 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
Load cell | 350 Ohm; -0.2..+3mV/V |
Pulse | Max. count 1000 Hz |
Frequency | 0.1..1000 Hz |
GENERATION FUNCTIONS | |
Current | 0.1..24 mA active and passive; protection ± 28 V |
Voltage (V) | 0.1..26 V |
Voltage (mV) | -10mV..+90mV |
Thermocouple | Type J,K ,T, E, N, R,S, B, L |
Thermoresistances (2 wires) | Pt100, Pt500, Pt1000, Cu50, Cu100, Ni100, Ni120 |
Load cell | 350 Ohm; -0.2..+3mV/V |
Pulse | Min 0.5 ms (1..24V) number of pulses can be set |
Frequency | 0.1..1000 Hz; resolution 50µs |
DATA LOGGER | |
Data logger | Yes |
Sampling time | >500 ms |
RAMP FUNCTION | |
Signal | Current/Voltage/TC/RTD/Load cell |
Functions | Single or Loop |
Type | Maximum 9 segments, ramp resolution 100ms, minimum ramp 1 second |